HP 8568B Manuel d'utilisateur Page 190

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 313
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 189
7. 3
MHz
Bandwidth Filter Adjustments
10 Hz Amplitude Adjustments
37.
38.
39.
40.
41.
42.
43.
Connect CAL OUTPUT to RF INPUT. Key in [INSTR PRESET),
[mj
9,
(jEEki--
10 Hz.
Adjust the instrument front panel FREQ ZERO control for
maximum signal amplitude on the CRT display.
Key in
CREs]
1 kHz and DISPLAY LINE
[ENTER).
Using the
DATA knob, place the display line at the signal trace.
Key in
[REs]
10 Hz.
Adjust the instrument front panel FREQ ZERO control for
maximum signal amplitude on the CRT display.
Adjust
A4A7R30
10 Hz AMPTD and
A4A7R41
10 Hz AMPTD equal
amounts to set the signal level the same as the reference level set
in step 39. See Figure 3-38 for location of 10 Hz AMPTD adjusts.
Repeat steps 37 through 42 until no further adjustment is
required.
3-76 Adjustments
Vue de la page 189
1 2 ... 185 186 187 188 189 190 191 192 193 194 195 ... 312 313

Commentaires sur ces manuels

Pas de commentaire