Chapter 11, Installing the Cellular Adapter
Connecting the Cellular Adapter to the HP 8920 Series Test Set
Installing the Cellular Adapter
I/Q Modulator and
Relative TX Power
Calibration
The I/Q Modulator and TX Power calibration routines must be run any time a
Cellular Adapter is disconnected and re-attached to a Test Set. TX Power should
also be recalibrated any time you suspect the relative TX Power readings are
drifting
.
Calibration Procedure
1. Press the TEST key to access the TEST (Main Menu) screen.
2. Under
Select Procedure Location:
choose
ROM
.
3. Under
Select Procedure Filename:
choose
PER_CALD
.
4. Press the k1 key to start the calibration program.
5. Follow the directions displayed on the screen.
a. Select and run the
I/Q Modulator
routine until it indicates that it has calibrated
at all frequencies; this will take several minutes. (Be sure to connect a short cable
from the DUPLEX OUT port to the ANT IN port before starting.)
b. Select and run the
Tx Power
routine. (Be sure to connect a short cable from the
DUPLEX OUT port to the RF IN/OUT port as instructed.)
6. When both calibration routines have been run, turn the Test Set off and back on to
enable the new calibration settings.
To Recalibrate Relative
TX Power
Repeat the previous procedure, skipping step 5a. Be sure to turn the power off and
back on to use the new power calibration settings before making more relative TX
Power measurements.