HP 83236B Manuel d'utilisateur Page 65

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 90
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 64
65
Chapter 4, Troubleshooting
Errors When Loading and Running the Software
Errors When Loading and Running the Software
Memory Overflow
Errors
Problem:
The random access memory (RAM) space of the Test Set is shared by IBASIC
programs and the Save/Recall storage registers. If you have saved a large number
of registers in your Test Set, you may encounter a “memory overflow” error when
you first attempt to load the Software.
Solution:
You will need to clear up some RAM space by deleting some or all of your saved
registers. You can do this selectively (one register at a time) or you can clear all
storage registers at once (globally).
NOTE: Clearing the registers, whether selectively or globally, is permanent. You cannot retrieve
the registers once deleted. You have an option to save registers to a RAM memory card, as
well. Consult the HP 8921 User’s Guide for instructions on this procedure.
To Clear a Single Register at a Time:
1 Press the RECALL key.
2 Scroll the cursor to the register to be deleted.
3 Press the ON/OFF key.
4 Press the YES key (same key) to confirm.
5 The register is cleared.
To Clear All Save Registers:
1 Press the RECALL key.
2 Scroll down to *Clr All* and press the knob.
3 Press the YES (ON/OFF) key to confirm.
4 All Save registers are now cleared.
Vue de la page 64
1 2 ... 60 61 62 63 64 65 66 67 68 69 70 ... 89 90

Commentaires sur ces manuels

Pas de commentaire