HP Saw E4419B Manuel d'utilisateur Page 187

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 204
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 186
HP4402.book : chapt5_2.fm 9 Fri Feb 12 15:32:45 1999
Specifications
Power Meter Supplemental Characteristics
HP E4419B User’s Guide 5-9
Settling Time
0 to 99% settled readings over the HP-IB.
For HP 8480 series power sensors
Manual filter, 10 dB decreasing power step:
Table 5-4: Settling Time
Auto filter, default resolution, 10 dB decreasing power step,
normal and x2 speed modes:
Number of
Averages
1 2 4 8 16 32 64 128 256 512 1024
Settling Time (s)
(Normal Mode)
0.15 0.2 0.3 0.5 1.1 1.9 3.4 6.6 13 27 57
Response Time (s)
(x2 mode)
0.15 0.18 0.22 0.35 0.55 1.1 1.9 3.5 6.9 14.5 33
10 dB
10 dB
10 dB
20 dB
Maximum dBm
Minimum dBm
Power
150 ms
6.6 s
500 ms
200 ms
Settling
Times
Sensor
Dynamic
Range
10 dB
10 dB
10 dB
20 dB
Maximum dBm
Minimum dBm
Power
Settling
Times
Sensor
Dynamic
Range
10 dB
10 dB
10 dB
20 dB
Maximum dBm
Minimum dBm
Power
Settling
Times
Sensor
Dynamic
Range
10 dB
10 dB
10 dB
20 dB
Power
150 ms
3.5 s
350 ms
180 ms
Settling
Times
Sensor
Dynamic
Range
10 dB
10 dB
10 dB
20 dB
Power
Settling
Times
Sensor
Dynamic
Range
10 dB
10 dB
10 dB
20 dB
Maximum dBm
Minimum dBm
Power
Settling
Times
Sensor
Dynamic
Range
Normal Mode x2 Mode
Typical
Typical
Vue de la page 186
1 2 ... 182 183 184 185 186 187 188 189 190 191 192 ... 203 204