Questionable Signal Status Register
Only bits 0 (VOLTage), 2 (TIME), and 8 (CALibration) are defined. All other bits
are always 0.
Bit 0 - VOLTage: Set (1) if an overload voltage is detected during a
measurement. Cleared (0) otherwise.
Bit 2 -TIME: Set (1) if the divide-by-n of the reference oscillator source can not
generate a sample rate that is within 1% of the rate specified by TRIG:TIMer1
or TRIG:TIMer2. Cleared (0) otherwise.
Bit 8 - CALibration: Set (1) if an error has been detected in the non-volatile
calibration memory. Cleared (0) otherwise.
Subsystem
Syntax
STATus
:OPC
:INITiate <
state
>
:OPERation|:QUEStionable
:CONDition? [query only]
:ENABle <
unmask
>
[:EVENt]? [query only]
:NTRansition <
unmask
>
:PTRansition <
unmask
>
:PRESet [no query]
:OPC:INITiate
STATus:OPC:INITiate <state > controls whether the *OPC, *OPC?, and *WAI
commands will complete immediately or whether they will wait for all
measurements to complete. With STATe OFF set, these commands will complete
immediately, which indicates that the parser is idle and ready for the next command,
even though the HP E1429 may still be in the INITiated state and taking a
measurement. With STATe ON set, these commands will wait for the Pending
Operation Flag set true by INITiate:IMMediate,
VINStrument:CONFigure:VME:MEMory:INITiate, or
VINStrument:CONFigure:LBUS:MEMory:INITiate to return false, indicating that
the trigger system is in the idle state and that all measurements have completed or
been aborted by the ABORt or *RST commands.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
state
boolean OFF|0|ON|1 none
STATus:OPC:INITiate
266 Command Reference STATus Subsystem