HP E1459A SCPI Command Reference 65
STATus:OPERation:PSUMmary[:EVENt]?
Returns the value of the Operation Status Port Summary Event Register as a signed
16 bit integer and then clears the register to 0.
Parameters None
Comments This is a destructive read so that all register bits are cleared after the read is
executed.
*RST does not affect the contents of the Status Operation Port Summary Event
Register.
*CLS clears the contents of the Status Operation Event Port Register.
STAT:PRESet does not affect the contents of the Status Operation Event Port
Summary register but does disable the reporting of the summary of this
register in bit 9 of the Status Operation Register.
STATus:PRESet
Presets the Status system registers and conditions.
Parameters None
Comments Resets the following registers and conditions:
Register Action Register Action
Status Byte none OPER Status condition none
Standard Event event none OPER Status event none
Standard Event enable presets to 0 OPER Status enable presets to 0
QUES Status Condition none OPER PSUM condition none
QUES Status Event none OPER PSUM event none
QUES Status enable presets to 0 OPER PSUM enable presets to 0
all transition filters none