HP 132A Manuel d'utilisateur Page 289

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 330
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 288
Chapter 4 Command Reference
Introduction
Programming Guide 4-123
*CLS
(Clear Status Command)
*CLS
Clears all event registers summarized in the status byte (Standard Event Status
Register, Operation Event Status Register, and Questionable Data Event Status
Register) and clears the Error Queue. The *CLS command will not clear data
memories or any other settings.
It also places the instrument in Operation Complete Idle Stateand “Operation
Complete Query Idle State(IEEE 488.2). This results in the disabling of any
prior *OPC command.
If *CLS immediately follows a program message terminator, the output queue and
the MAV bit are cleared because any” new program message after a program
message terminator clears the output queue.
This command will clear any displayed error message from the front panel.
Vue de la page 288
1 2 ... 284 285 286 287 288 289 290 291 292 293 294 ... 329 330

Commentaires sur ces manuels

Pas de commentaire