Vue de la page 174
165
Selecting and Using "Other" Features
Running Self Test
4 Press toactivatethe SelfTest. TESTSTATUS RUNNING
will be displayed.
The information pertaining to TEST TYPE, TEST NUMBER and
SUBTEST NUMBER will change as the Self Test progresses.
Ifthe HP 37718Ais functioning correctly,aftera time of2 to 3minutes,
TEST STATUS PASSED is displayed.
If TEST STATUS [FAIL nnn] is displayed, the HP 37718A should be
returned to a service office for repair.
NOTE Each individual self test requires unique loopback connections. To obtain
a list of the connections required move the highlighted cursor to
CABLING INFO and press . The Loopbacks list will appear on the
display.
RUN/STOP
SET
Vue de la page 174