84
Making Measurements
Measuring Jitter
Measuring Jitter
Description: Jitter and error measurements are made simultaneously when a jitter
option is fitted. The measurements are made on the normal input to the
DSn or SONET receiver and the interface selections are the normal DSn
or SONET Receiver selections.
Cumulative and Short Term results of Jitter Amplitude and Jitter Hits
are provided on the display.
Graph and Text results for Jitter Transfer and Jitter Tolerance are also
provided.
HOW TO: 1 If measuring Jitter on a DSn signal, set up the receive DSn interface
and the receive Jitter interface. See “Setting DSn Receive Interface”
page 15 and “Setting Jitter Receive Interface” page 18.
2 If measuring Jitter on an SONET signal, set up the receive SONET
interfaceand the receiveJitter interface.See “SettingSONET Receive
Interface” page 17 and “Setting Jitter Receive Interface” page 18.
RESULTS
JITTER
Vue de la page 93