HP Waterskis E1312A Manuel d'utilisateur Page 84

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 212
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 83
84 Multimeter Command Reference Chapter 3
CALibration?
CALibration? performs a calibration using the specified calibration value set by the
CALibration:VALue command and queries the calibration response to verify a
successful calibration.
Comments Execution of this command begins the electronic adjustment for the function
and range the multimeter is set to. The adjustment is performed based on the
value stated in the
CAL:VAL command and the multimeter expects that value at
the input terminals.
The command returns “0 to indicate there are no calibration errors and the
calibration was performed. A “
1” is returned if a calibration error occurs and a
calibration is unable to be performed. The error message is reported to the
output buffer.
You must set CALibration:SECure:STATe OFF <code> to allow a calibration
to be performed. This requires that you know the calibration secure code. The
secure state enabled prevents unauthorized calibration of the multimeter.
Example Calibrate the Active Function and Range Using the CAL:VALue
CAL?
Perform the calibration.
monitor the status byte to detect calibration operation complete
enter statement
Enter cal response into computer to verify the
calibration was successful.
Vue de la page 83
1 2 ... 79 80 81 82 83 84 85 86 87 88 89 ... 211 212

Commentaires sur ces manuels

Pas de commentaire