HP Waterskis E1312A Manuel d'utilisateur Page 97

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 212
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 96
Multimeter Command Reference 97Chapter 3
DATA
The multimeter can store up to 512 readings in internal memory. The DATA
command allows you to determine how many readings are currently stored.
Subsystem Syntax DATA
:POINts?
:POINts?
The INITiate command uses internal memory to store readings prior to a FETCh?
command e.g., when a measurement is initiated by the
INITiate command. You can
query the number of stored readings in memory by sending the
DATA:POINts?
command.
Comments INITiate command uses internal memory to store readings prior to using a
FETCh? command. You use the DATA:POINts? command to query the
number of readings stored in internal memory to determine the amount of data
space to allocate on your computer to receive the data.
Vue de la page 96
1 2 ... 92 93 94 95 96 97 98 99 100 101 102 ... 211 212

Commentaires sur ces manuels

Pas de commentaire