HP E6380-90019 Manuel d'utilisateur Page 139

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 228
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 138
Chapter 5 139
Analog Measurements - Screens and Control Fields
Marker To Peak+ Marker To (spectrum analyzer)
Marker To Peak+
Peak+ causes the marker to move to the maximum value of the average
level measured on the display.
Because this functions look at the average value for each displayed
pixel, the marker may not appear directly on the displayed peak of a
noisy signal.
Screens on which this field is displayed
SCOPE: marker controls menu
Marker To Peak-
Peak causes the marker to move to the minimum value of the average
level measured on the display.
Because this functions look at the average value for each displayed
pixel, the marker may not appear directly on the displayed peak of a
noisy signal.
Screens on which this field is displayed
SCOPE: marker controls menu
Marker To (spectrum analyzer)
These two fields position the marker as follows.
Peak moves the marker to the highest peak and enters the location
in the Position field.
Next Peak moves the marker to the next peak to the right and
enters the location in the Position field.
Center Freq changes the center frequency value to match the
current position of the marker.
Ref Level changes the reference level setting to match the current
position of the marker.
Screens on which this field is displayed
SPEC ANL: marker controls menu
Main Menu
Vue de la page 138
1 2 ... 134 135 136 137 138 139 140 141 142 143 144 ... 227 228

Commentaires sur ces manuels

Pas de commentaire