Chapter 2 41
Operating Overview
Calibrating the Channel Power Measurement
Calibrating the Channel Power Measurement
To calibrate a channel power measurement at the RF IN/OUT port
using the internal CDMA generator
Step 1. Go to the CDMA ANALYZER screen and choose the Chan Pwr
measurement.
Step 2. Set the Tune Freq field (or RF Channel field) to the center
frequency of your CDMA signal.
Step 3. Select the Chn Pwr Cal field and wait until the calibration is
complete.
Main Menu