96 Chapter 4
CDMA Measurements - Screens and Control Fields
Car FT Chn Pwr Cal
Car FT
This field is a measurement display field. See "Carrier Feedthrough" on
page 87 for more information about this measurement.
Screens on which this field is present
COD DOM (Code Domain Analyzer)
Chn Pwr Cal
This field is displayed when the channel power measurement is
selected. It is used to calibrate the channel power measurement. When
this field is selected, an internally generated calibration signal is
measured using the average power technique. Channel power is also
measured and a correction factor is generated. This correction factor is
applied to subsequent channel power measurements. Calibration
should be performed whenever a new set of measurements is made and
whenever the frequency of the measured signal is changed. See
"Calibrating the Channel Power Measurement" on page 41 for more
information.
Screens on which this field is present
CDMA ANALYZER
Main Menu