HP Data Explorer 4 Series Manuel d'utilisateur Page 180

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 447
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 179
Chapter 4 Examining Chromatogram Data
4-8 Applied Biosystems
4
From the
Spectrum window
To create an extracted ion chromatogram for a mass range
from the Spectrum window:
1. Click the Chromatogram window to activate it.
2. Select Duplicate Active Trace from the Display menu to
keep the original data displayed after processing.
3. In the Spectrum window, right-click-drag over the mass
region of interest in the extracted ion chromatogram. The
width of the box you draw defines the precise mass range
used in the extracted ion chromatogram.
The extracted ion chromatogram is displayed in the
Chromatogram window (Figure 4-2) with the mass
range indicated in the trace label.
Figure 4-2 Extracted Ion Chromatogram
4. To return to the original trace, see Returning to the
original trace on page 4-4.
Extracted ion chromatogram
(center window)
Vue de la page 179
1 2 ... 175 176 177 178 179 180 181 182 183 184 185 ... 446 447

Commentaires sur ces manuels

Pas de commentaire