HP 8591C Manuel d'utilisateur Page 218

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 550
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 217
I
-
Making Measurements Using Spectrum Analyzer Mode
Stimulus-Response Measurements
5. Since we are only interested in the rejection of the low-pass
Iilter,
tune
the spectrum analyzer’s center frequency so that the roll-off of the filter
comprises the majority of the trace on the display (see Figure 5-6).
6. Decrease the resolution bandwidth to increase sensitivity, and narrow
the video bandwidth to smooth the noise. In Figure 5-7, the resolution
bandwidth has been decreased to 10
kHz.
REF
48.8
dBmV
AT
10
dB
PES
PEAK
AUTO
ML\N
LOG
10
dB/
“IO
BW
a
MAN
VBW/RBW
RATIO
VI0
AVG
ON
02
WA
SB
SC
FC
\
CORR
\
‘4
EM1
BW
W”U
I
CENTER
350
.O
MHZ
SPAN
500.0
MHz
#RES
BW
10
kHr
VBW
10
kHz
SWP
15.0
set
Figure
5.7.
Decrease the Resolution Bandwidth to Improve Sensitivity
5-18
Vue de la page 217
1 2 ... 213 214 215 216 217 218 219 220 221 222 223 ... 549 550

Commentaires sur ces manuels

Pas de commentaire