I
-
Spectrum Analyzer Mode Key Descriptions
Spectrum Analyzer Mode Functions
CLEAR
WRITE B
erases any data previously stored in trace B and
continuously displays any signals detected during the sweep
of the analvzer. This function is activated at power on and
by pressing
CPRESET].
Changing the trace mode of trace C to clear write or
minimum hold can change the trace mode of trace B. If trace
B is in clear-write mode or maximum-hold mode when trace
C is changed to clear write or minimum hold, the trace
mode of trace B is changed to store blank.
The following table shows the trace mode of trace B before
and after changing trace C to clear-write or minimum-hold
trace mode.
Changing the trace mode of trace B to clear write or
maximum hold can change the trace mode of trace C. If
trace C is in clear-write mode when trace B is changed to
clear write or minimum hold, the trace mode of trace C is
changed to minimum hold.
Trace Mode of Trace C Trace Mode of Trace C
11
Front-Panel Key Access
(m)