HP 132A Manuel d'utilisateur Page 243

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 330
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 242
Chapter 4 Command Reference
Introduction
Programming Guide 4-77
How to Use the Measurement Instruction Commands (Cont.)
Use
:SENS:EVEN:LEV 0V
to set the trigger level to 0 Volts.
Use
:INITIATE
to start the measurement.
Use
:FETCH?
to query for result.
Firmware Revision Work-Around Commands
The following applies to HP 53131A instruments with Firmware Revision 3413:
The three commands listed below should NOT be used.
:MEASure[:SCALar][:VOLTage]:MAXimum? <parameters>[<source_list>]
:MEASure[:SCALar][:VOLTage]:MINimum? <parameters>[<source_list>]
:MEASure[:SCALar][:VOLTage]:PTPeak? <parameters>[<source_list>]
The work-around commands which should be substituted are listed below.
:CONFigure[:SCALar][:VOLTage]:MAXimum <parameters>[<source_list>]
:INIT; *WAI; FETCh?
:CONFigure[:SCALar][:VOLTage]:MINimum <parameters>[<source_list>]
:INIT; *WAI; FETCh?
:CONFigure[:SCALar][:VOLTage]:PTPeak <parameters>[<source_list>]
:INIT; *WAI; FETCh?
The three commands listed below should NOT be used.
:READ[:SCALar][:VOLTage]:MAXimum?
:READ[:SCALar][:VOLTage]:MINimum?
:READ[:SCALar][:VOLTage]:PTPeak?
Vue de la page 242
1 2 ... 238 239 240 241 242 243 244 245 246 247 248 ... 329 330

Commentaires sur ces manuels

Pas de commentaire