Chapter 4 Command Reference
Introduction
Programming Guide 4-137
*RST
(Reset Command)
*RST
This event command performs an instrument reset .
The reset performs the following:
sets instrument settings to their *RST states,
disables macros,
places instrument in “Operation Complete Idle Stateand “Operation
Complete Query Idle State,and
clears (empties) the Key Queue
The reset does not affect:
the macros defined with *DMC,
the calibration data,
the Service Register Enable or the Standard Event Status Enable,
the Output Queue, and
the IEEE 488.1 address or the state of the IEEE 488.1 interface.
See the section titled “*RST Response,page 2-36, in Chapter 2 of this guide for
a complete listing of the *RST state.
Each command description in this chapter (Chapter 4) includes the *RST state in
the “Commentportion of the definition.