ARM
The ARM subsystem selects the number of scanning cycles (1 - 32767) for
each
INIT command. This command does not apply to downloaded scan
lists. For downloaded scan lists, use
INIT:CONT ON for multiple scans.
Subsystem Syntax ARM
:COUNt <
number> MIN | MAX
:COUNt? [MIN | MAX]
:COUNt ARM:COUNt <number> MIN | MAX allows scanning cycles to occur a
multiple of times (1 to 32767) with one
INITiate command and when
INITiate:CONTinuous OFF | 0 is set. MIN sets 1 scanning cycle and MAX
sets 32,767 scanning cycles.
Parameters
Parameter Name Parameter Type Range of Values
number
Numeric 1 - 32767 | MIN | MAX
Comments Number of Scans: Use only numeric values between 1 to 32767 for
the number of scanning cycles (default is 1).
Related Commands: ABORt, INITiate[:IMMediate]
*RST Condition: ARM:COUNt 1
Example Setting Ten Scanning Cycles
ARM:COUN 10 Set 10 scanning cycles.
SCAN (@100:115) Sets channel list.
INIT Starts scanning cycle.
:COUNt? ARM:COUNt? [MIN | MAX] returns the current number of scanning cycles
set by
ARM:COUNt. ARM:COUN? MIN returns 1 and ARM:COUN? MAX
returns 32767.
Parameters
Parameter Name Parameter Type Range of Values
MIN | MAX Numeric MIN = 1, MAX = 32767
Comments Related Commands: INITiate[:IMMediate]
Example Query Number of Scanning Cycles
ARM:COUN 10 Sets 10 scanning cycles.
ARM COUN? Query number of scanning cycles;
returned value is 10.